Difference between revisions of "20191206 FARO Arm - Movement Filter set to zero"

From ATTWiki
Jump to: navigation, search
(OUR DISCOVERY)
(OUR DISCOVERY)
Line 12: Line 12:
 
==OUR DISCOVERY==
 
==OUR DISCOVERY==
 
The FARO driver has a new Movement Filter option that has an impact on scanning whether or not is enabled.
 
The FARO driver has a new Movement Filter option that has an impact on scanning whether or not is enabled.
 +
<br><br>
 +
In this screen image of the Probe Management window, we show that even with the Movement Filter disabled, it is critical to have a non-zero value in the Movement Filter box on the right.  In this case, we enabled the option, then slid the slider all the way to the right, then disabled the option.
  
 
[[image:FARO_MFilter_001.png|700px]]
 
[[image:FARO_MFilter_001.png|700px]]

Revision as of 17:07, 6 December 2019

Back to VTube-LASER

Contents

CUSTOMER ARM MODEL

  • FARO Arm
  • Edge
  • BluLaser Scanner

CUSTOMER ISSUE

The customer could not get any laser data into VTube-LASER.

OUR DISCOVERY

The FARO driver has a new Movement Filter option that has an impact on scanning whether or not is enabled.

In this screen image of the Probe Management window, we show that even with the Movement Filter disabled, it is critical to have a non-zero value in the Movement Filter box on the right. In this case, we enabled the option, then slid the slider all the way to the right, then disabled the option.

FARO MFilter 001.png

OTHER ACTIONS

The customer calibrated the ball probe (it passed spec) and then calibrated the laser probe. It did not pass spec. However, it did calibrate to 0.0675 mm (0.0027") - which is very good for tube fabrication. The customer saved the calibration results anyway and continued.

Since the probes were recalibrated, it was also necessary to re-establish the CUT PLANE.

RESULT

VTube-LASER now scans the 5 mm wire with no problems.