'''Set the Scan Rates to 1/1.'''<br>
+
'''Set the Scan Rate to 1/1.'''<br>
 
* This value will allow the laser stripes to come into VTube-LASER as quickly as the driver can send them.<br><br>
 
* This value will allow the laser stripes to come into VTube-LASER as quickly as the driver can send them.<br><br>
  

Revision as of 19:27, 25 December 2015

Vtube-laser logo 1.96.png

Vtl screen hd scanner without logo.png



Ideal Setup for HD Scanner in VTube-LASER

The HD blue line scanner has increased capabilities in point density that causes us to recommend some unique settings in FARO scanner setup. These settings work very well in VTube-LASER.

You must be connected to a FARO arm to perform these steps.

STEP 1

In VTube-LASER window click on Scanner Algorithm label above the viewport window.

The alternative way to get into the FARO LLP Scanner window is to click on Measure Setup in the Navigation Pane > Arm tab > Scanner Control button.

Vtube-laser v2.2 leftmouseclick scanneralgorithm label.png
Vtube-laser v2.2 leftmouseclick scanneralgorithm label closeup.png

STEP 2

Set the Scan Rate to 1/1.

  • This value will allow the laser stripes to come into VTube-LASER as quickly as the driver can send them.

Set the Scan Density to 1/16

  • This value will reduce density of the points in the laser line returned to VTube-LASER.

Since VTube-LASER removes some of points with a Maximum Number of Points Allowed in Cylinder Laser Stripe filter, it works faster if the FARO driver removes some of those points in advance. Both scanning and calculation can be many times faster when the Scan Density is reduced to anything less than 1/1.

Vtube-laser v2.2 farol llp scan rate density.png

STEP 3

For most materials set the Algorithm to Automatic - Normal.

(For chrome and dark material colors, it can be helpful to switch to Automatic - HDR.)

Vtube-laser v2.2 farol llp scan algorithm.png

STEP 4

Press the Close button in the FARO Arm Utilities window.

Vtube-laser v2.2 armutilities dialog press close.png

STEP 5

Observe that the new ball probe selection will appear in the Arm grid.

Vtube-laser 2.2 measuresetup menu 6mmball.png