20191206 FARO Arm - Movement Filter set to zero

From ATTWiki
Revision as of 17:12, 6 December 2019 by Mcone (Talk | contribs)

Jump to: navigation, search

Back to VTube-LASER

Contents

CUSTOMER ARM MODEL

  • FARO Arm
  • Edge
  • BluLaser Scanner

CUSTOMER ISSUE

The customer could not get any laser data into VTube-LASER.

OUR DISCOVERY

The FARO driver has a new Movement Filter option that has an impact on scanning whether or not is enabled.

FARO MFilter 003.png

SOLUTION

It is critical to have a non-zero value in the Movement Filter box on the right. In this case, we enabled the option, then slid the slider all the way to the right, then disabled the option.

FARO MFilter 001.png

RESULT

VTube-LASER now scans the 5 mm wire with no problems.