Difference between revisions of "20190625 HEXAGON Arm - Could not scan a tube and get good results"

From ATTWiki
Jump to: navigation, search
(Created page with "Back to VTube-LASER ==CUSTOMER ARM MODEL== *HEXAGON Arm *Model 7325 ==OPERATOR ISSUE== Scanning was poor and failed often. ==OUR DISCOVERY== *MANUAL scanning exposure w...")
 
(OUR DISCOVERY)
 
Line 9: Line 9:
  
 
==OUR DISCOVERY==
 
==OUR DISCOVERY==
*MANUAL scanning exposure was being used in RDS.
+
*The exposure mode was set to MANUAL inside RDS.
 
+
  
 
==OTHER ACTIONS==
 
==OTHER ACTIONS==

Latest revision as of 15:29, 27 June 2019

Back to VTube-LASER

Contents

CUSTOMER ARM MODEL

  • HEXAGON Arm
  • Model 7325

OPERATOR ISSUE

Scanning was poor and failed often.

OUR DISCOVERY

  • The exposure mode was set to MANUAL inside RDS.

OTHER ACTIONS

  • The customer changed the exposure mode to Automatic.


RESULT

The system measured perfectly.