20190426 HEXAGON Arm - Slow Scanning - We Found Incorrect Scanning Profile

From ATTWiki
Revision as of 20:23, 26 April 2019 by Mcone (Talk | contribs)

(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search

Back to VTube-LASER

Contents

CUSTOMER ARM MODEL

  • HEXAGON Arm
  • Model RA8325

OPERATOR ISSUE

The scanning seemed to have delays.

OUR DISCOVERY

  1. The wrong scanning profile was being used in RDS.
  2. The Point Sampling was set to 100% - which is way too dense for VTube-LASER.

OTHER ACTIONS

  1. We opened RDS and checked the current Scanning Profile.
  2. We set the scan settings point sampling to 25%.
  3. We made sure that this was associated with and saved to a new profile named "VTube".
  4. We set the active profile to VTube.

RESULT

VTube-LASER measures quickly and perfectly